Laminar Flow Hoods, also known as Clean Benches, serve a crucial role in safeguarding product integrity in highly sensitive environments by preventing contamination.
Inspection
Calibration
Validation
Repair
In adherence to industry regulations, Laminar Flow Hoods are required to pass certification assessments during installation and bi-annually. Furthermore, recertification must be administered following HEPA filter replacement, repairs, or relocation incidents.
This test is done to ensure that the airflow velocity within the cabinet meets the required standards specified by the manufacturer.
This test verifies that the number of particles inside the cabinet is within acceptable limits and that the HEPA filter is capturing the majority of them.
This test is done to ensure that there is no leakage in the cabinet's HEPA filter, which could potentially result in contamination.
This test is done to visualize the airflow pattern within the cabinet using smoke, which helps in identifying if there are any areas with poor air circulation.
This test verifies the effectiveness of the alarms and evaluates the proper setup of the units.
This test is done to ensure that the lighting inside the cabinet is sufficient for the operator to accurately perform their work.
This test is done to ensure that there are no vibrations that may interfere with the operation of sensitive equipment.
This test is done to ensure that the cabinet operates at a safe noise level for laboratory personnel.
This test ensures the proper functioning of the UV lights.
This test ensures that the electrical components are in good working order and pose no danger of electrical hazards.
Validation Certificate
E-Service Report
Service Sticker
Horizontal Laminar Flow Cabinet
Vertical Laminar Flow Cabinet